K88800 Heat of Combustion of Liquid Hydrocarbon Fuels
by Bomb Calorimeter

Test Method
Heat of combustion is determined in this test method by burning a weighed sample in an oxygen bomb calorimeter under controlled conditions. The heat of combustion is computed from temperature observations before, during and after combustion with proper allowances for thermochemical and heat transfer corrections. Either isothermal or adiabatic calorimeter jackets can be used.


Click on the image to view a larger photo of the instrument.

K88800 Automatic Calorimeter
The automatic calorimeter is the latest system for determining gross calorific values of liquids and solid fuels. A higher level of automation with extremely simple handling characterizes this device.

In addition to the Isoperibolic measurement procedure, a Dynamic (reduced-time) mode is also available for the user. Different working temperatures can be selected for both procedures based on the temperature of the connected water.

To provide a supply of cooling water, the calorimeter can be connected to a standard thermostat or an appropriate permanently installed water connection, with a connection valve. The unit is equipped with a very convenient operating panel through which operation of the device takes place. The graphical display with active back lighting displays the appropriate status messages. The temporal course of a measurement that has been started and all current parameters of the weighed in sample can be constantly monitored and are arranged to be clearly visible.

Connections for analysis scale, printer, sample rack for identifying and managing samples are already integrated into the basic device. The network connection and the special configuration for data exchange can be implemented at any time with LIMS.

In combination with special halogen-resistant decomposition vessels quantitative decompositions can be performed to determine halogen and sulfur content.

Specifications
Conforms to the specifications of:
ASTM D240; D4809; D5865; D1989; D5468; E711; ISO 1928;
DIN 51900; BS1016
Measurement range: 40,000 J
Measuring mode: Isoperibolic 25°C; Isoperibolic 30°C;
Dynamic 25°C; Dynamic 30°C
Isoperibolic Measuring Time: Approximately 22 min
Dynamic Measuring Time: Approximately 7 min
Oxygen Operating Pressure: 30 bar
Cooling Medium: Water via line, flow through quantity 60 + 10 liters / hour
Water Operating Pressure: 1 – 1.5 bar max.
Water Test Pressure: 10 bar
Interfaces: Serial (RS232); Parallel; Keyboard; Sample rack; External monitor


Dimensions lxwxh,in.(cm)
440 x 450 x 500 (WxDxHmm)
Net Weight: 30Kg


Ordering Information
Catalog
Number
Product Description
Electrical
Requirements
K88800 Automatic Calorimeter
115V 60Hz
K88890 Automatic Calorimeter
220V 50Hz

Accessories
Catalog
Number
Product Description
K88800-1
Cooling water supply unit, 115V 60Hz
K88890-1
Cooling water supply unit, 220V 50Hz
K88800-2
Pressure Gauge, Oxygen To reduce the pressure of the oxygen cylinder to 30 bar
K88800-3
Standard Decomposition Vessel
K88800-4
Decomposition Vessel, Halogen Resistant For quantitative decomposition determine halogen and sulfur content
K88800-5
Connection valve Required for permanently installed water connection